Fluke ESA715 Electrical Safety Analyzer
Ensuring Electrical Safety with Confidence
Product Overview
Fluke ESA715 Electrical Safety Analyzer Price in Bangladesh
Fluke ESA715 Electrical Safety Analyzer is a professional biomedical testing instrument, and healthcare facilities searching for the Fluke ESA715 Electrical Safety Analyzer Price in Bangladesh often require a dependable solution for performing comprehensive electrical safety inspections. Designed for medical equipment testing and preventive maintenance, this analyzer helps ensure that healthcare devices operate safely and meet required performance standards.
Built with advanced testing capabilities, the Fluke ESA715 Electrical Safety Analyzer enables technicians and biomedical engineers to perform accurate electrical safety evaluations with efficiency and confidence. It supports a wide range of testing procedures, helping identify potential electrical hazards and ensuring that critical medical equipment remains safe for patient use. This contributes to improved operational reliability and healthcare safety.
The analyzer is designed with a user friendly interface that simplifies testing workflows and reduces the time required for inspections. Its intuitive operation allows maintenance professionals to perform safety assessments efficiently while maintaining consistent and reliable results. This improves productivity and supports effective equipment management programs within healthcare environments.
Portability is another important advantage of the Fluke ESA715 Electrical Safety Analyzer. Its compact and lightweight construction allows technicians to carry the device easily between departments, facilities, and service locations. This flexibility makes it suitable for both in house biomedical maintenance and field service applications.
Durability is a key feature of the Fluke ESA715 Electrical Safety Analyzer. Engineered for regular use in demanding healthcare and technical environments, it delivers dependable performance over extended periods. Its robust construction ensures long term reliability and helps maintain testing accuracy throughout continuous operation.
Routine electrical safety testing is essential for reducing equipment related risks and maintaining compliance with healthcare safety requirements. The Fluke ESA715 Electrical Safety Analyzer helps organizations detect potential issues before they affect clinical operations, supporting proactive maintenance strategies and improving equipment dependability.
Combining accuracy, durability, portability, and ease of use, the Fluke ESA715 Electrical Safety Analyzer provides a reliable solution for medical equipment safety verification. Its professional capabilities make it an excellent choice for hospitals, clinics, biomedical departments, and service professionals committed to maintaining high standards of electrical safety and equipment performance.
Specifications
Specifications: Fluke ESA715 Electrical Safety Analyzer
Safety standard compliance IEC 61010-1: Overvoltage category II, pollution degree 2
IEC 61010-2-034: Measurement CAT II 300 V
Measurement equipment
standard compliance
IEC 61557-16:2014, except IP40 for certain equipment outlets
Dimensions (w x d x h) 214 x 207 x 92 mm (8.4 x 8.1 x 3.6 inches)
Weight 1.3 kg (2.7 pounds)
Ingress protection IP40 per IEC 60529, excluding equipment outlet
Operating temperature 0 to +35 °C (+50 to +95 °F)
Operating humidity 10-90 %, non-condensing
Storage temperature -20 to +60 °C (-4 to+ 140 °F)
Storage humidity 5-95 %, non-condensing
Battery charging temperature 8 to +28°C (+46 to +82°F)
Altitude 100-127 V ac mains voltage and ≤5 150 V on input jacks: ≤ 5000 m
200-240 V ac mains voltage and ≤5 300 V on input jacks: ≤ 2000 m
Battery Built-in rechargeable lithium-ion (<3600 mAh, 3.6 V)
Battery life Up to 2 h
Connectivity 1 x USB-C for PC communication, 2 x USB-A for peripherals
Display 5-inch touchscreen
Data storage > 10 000 measurements
Power (region dependent) 90-132 V ac, 20 A MAX, 47-63 Hz
90-132 V ac & 180-264 Vac, 10 A MAX, 47-63 Hz
90-132 V ac & 180-264 V ac, 15 A MAX, 47-63 Hz
90-132 V ac & 180-264 V ac, 16 A MAX, 47-63 Hz
CSA approval
US version,
90-132 V ac, 20 A MAX, 47-63 Hz
CSA approved
NEMA 6-15 version,
90-132 V ac & 180-264 V ac,
15 A MAX, 47-63 Hz
CSA approved
Measurement Specifications
Mains voltage
Range 0-264 V ac rms, limited by power specification
Accuracy ±(2 % + 0.2 V)
Point to point voltage
Range 0-300 V ac rms, limited by altitude specification
Accuracy ±(2 % + 0.2 V)
Protective earth resistance & Point to point resistance
Range 0-20 Ω
Accuracy ±(1 % + 0.01 Ω) at ≤ 2 Ω
±(1 % + 0.1 Ω) at >2 Ω
Test current Square wave of at least ±200 mA for ≤ 2 Ω
Open circuit voltage Max ±24 V
Equipment Current
Range 0-20 A ac rms, limited by power specification
Accuracy ±(5 % + 0.05 A)
Max duty cycle 0-10 A: continuous
10-15 A: 7 min on/3 min off
15-20 A: 5 min on/5 min off
Note: Equipment outlet will be turned off if max duty cycle is exceeded.
Leakage Current
Modes ac + dc (true rms), ac only, dc only
Mains on applied part leakage test: ac only
Patient load AAMI ES1:1993
IEC 60601-1:2005 / IEC 62353:2014
See users manual for more information.
Crest factor ≤2
Range 0 μA - 20 mA
Accuracy dc & 20 Hz - 0.5 kHz: ±(1 % + 1 μA)
0.5 kHz - 50 kHz: ±(2.5 % + 1 μA)
50 kHz - 1 MHz: ±(5 % + 1 μA)
Mains on applied part (Applies to: mains on applied part leakage, direct applied part leakage, alternative applied part leakage, alternative
equipment leakage)
Test voltage Mains voltage ±5 %
Current limit 1 mA ± 25 % at 115 V for AAMI ES1, NFPA 99
3.5 mA ± 25 % at 230 V for IEC 60601-1, IEC 62353 and EN 50678 / EN 50699
7.5 mA ± 25 % at 230 V for AS/NZS 3551
Additional uncertainty Up to ±2 μA for 120 V, up to ±4 μA for 230 V
Insulation Resistance
Test voltage 250 V dc
and 500 V dc
Range 0.1-100 MΩ
Accuracy ±(2 % + 0.2 MΩ) at ≤ 10 MΩ
±(7.5 % + 0.2 MΩ) at >10 MΩ
Test voltage accuracy +20 % / -0 %
Test voltage 50 V dc and
100 V dc
Range 0.1-20 MΩ
Accuracy ±(10 % + 0.2 MΩ)
Test voltage accuracy +30 % / -0 %
Short-circuit current 2 mA ± 0.25 mA
Maximum load capacitance 2 μF
ECG simulation
Frequency accuracy ±2 %
Amplitude accuracy ±5 %, for 2 Hz square wave
Waveforms ECG complex 30, 60, 120, 180 and 240 BPM
Square wave 0.125 and 2 Hz, 50 % duty cycle
Triangle wave 2 Hz
Pulse, 63 ms 30 and 60 BPM
Ventricular fibrillation
Respiration simulation
Rate Apnea (0 BrPM) and 10-100 BrPM in 10 BrPM steps
Waveforms Normal
Inspiration : expiration ratio 1:1
Impedance baseline 1000 Ω ± 5 % between leads
Impedance variation 1 ± 0.15 Ω
Respiration lead LL or LA, user selectable
Features
Key Features: Fluke ESA715 Electrical Safety Analyzer
• 5” Touchscreen with intuitive user interface
• Small, portable, ergonomic design with an integrated handle
• OneQA-enabled workflow automation eliminates the need for a laptop
• Wireless capability allows you to run test procedures in the field and automatically sync results
• Ten applied parts posts with innovative LED lighting guidance (ECG & Banana)
• Insulation resistance from 50V, 100V, 250V and 500V
• On-board automation with automated test sequences for compliance to key electrical safety standards NFPA 99 / AAMI ES1, AS/NZS 3551, IEC 62353, and parts of EN 50678 / EN 50699, and IEC 60601-1
• ECG waveform tests and dual-lead measurements combine the functionality of a simulator and safety analyzer in a single test tool
• Ability to upload pictures, instructions, and text to procedures
• Easy data entry through barcode scanner, external keyboard, or on-screen keyboard
• 20 A at 120 V equipment current capability
• Replaceable mains fuses keep your unit in the field and out of the repair shop
• Rigorously tested for rugged field applications with CE and CSA in addition to Fluke-quality ruggedness for long-term reliability
• Enjoy precision, compliance, and total peace of mind through our comprehensive service and coverage program, eliminating unexpected costs with our Protect+ Program
• Global support network delivering prompt service and peace of mind to Fluke Biomedical customers worldwide